How to Pre-Approve Medical AI Updates: Unpacking the TGA's New Change Control Rules
Australia's Therapeutic Goods Administration has released technical guidance on Predetermined Change Control Plans for medical device software. Listeners will learn how manufacturers can get pre-market approval for future AI model retraining and software updates without filing repeated submissions, provided they set strict, pre-defined boundaries.
Key points
- Predetermined Change Control Plans allow medical device software to be updated iteratively without repeated pre-market assessments if changes are pre-specified and bounded.
- The guidance specifically targets higher-risk software like AI and machine learning that requires model re-training, excluding exempt or Class I software unless it has a measuring function.
- Changes to the original intended purpose or broad concepts like "future improvements" cannot be included in a plan.
- Data used for retraining must be justified for its relevance to the Australian population, requiring bias and equity assessments.
- The rules apply strictly to software; hardware changes still require traditional regulatory pathways.
Source: TGA Approach to Predetermined Change Control Plans - Therapeutic Goods Administration, 2026
This spot is available. Reach clinicians, health-system leaders and medtech and pharma teams following AI in medicine. Sponsor the show
This episode is an AI-generated conversation summarising a public document; the hosts' voices are synthetic. It is for information only and is not medical advice. Always refer to the original source.
Transcript
Maya: Artificial intelligence models are designed to learn and evolve over time, but historically, medical device regulators demand that every single software update be frozen in time and individually approved.
Sam: That paradigm changes right now. Today we are looking at how Australia's regulator is laying out a fast lane to pre-approve AI updates and medical software changes before they even happen.
Maya: Before we dive in, a quick note that our voices are AI-generated, and this episode is a summary of a public document. We are unpacking the technical guidance document called TGA Approach to Predetermined Change Control Plans, published by the Therapeutic Goods Administration.
Sam: This is a massive shift for anyone building or buying medical software. We are talking about Predetermined Change Control Plans. Basically, instead of asking for permission every single time you tweak an algorithm or update an interface, you get the actual plan for making future changes approved upfront by the regulator.
Maya: Exactly. The Therapeutic Goods Administration, known as the TGA, helped develop the 2025 International Medical Device Regulators Forum draft guidance on this topic. This specific technical guidance aligns with that international document. It is specifically intended for manufacturers, developers, and sponsors of medical device software.
Sam: And when they say medical device software, they mean software as a medical device, software embedded in other devices, and in vitro diagnostic software. But the document calls out that this framework is particularly relevant for artificial intelligence or machine learning enabled software that gets updated iteratively, or where the algorithm model is actually re-trained.
Maya: Right. Because if you do not have one of these predetermined plans in place, making those changes would normally require regulatory authorisation prior to implementation each time any of the changes were made. That creates a huge bottleneck for innovation and delays getting better tools to clinicians.
Sam: So how does a Predetermined Change Control Plan actually work? If I am a medical software developer, what exactly am I handing over to the regulator?
Maya: You are giving them a highly detailed roadmap. The document states a plan describes 3 things. First, you need a description of the specific planned changes to the medical device software. Second, you must provide the change plan or protocol for implementing and controlling those changes, which has to include predefined acceptance or performance criteria. And 3, an assessment of impacts from those changes.
Sam: Okay, so I tell them exactly what I want to change, exactly how I will test that change to prove it works safely, and what the ripple effects on the broader system might be. If the TGA approves that roadmap, I can just execute the plan later without waiting in line for another formal assessment.
Maya: Precisely. But there is a very strict fence around what is allowed. These plans are most suitable for predictable, well-characterised changes. You cannot just ask for a blank check to modify your software however you see fit in the future.
Sam: Let us talk about that regulatory fence. What kinds of changes are actually allowed inside one of these predetermined plans?
Maya: The document gives several explicit examples of what is acceptable. The big one is pre-specified artificial intelligence or machine learning model re-training. But it has to be within defined data characteristics, performance floors, drift monitoring triggers, and acceptance criteria.
Sam: So a developer can say, we are going to retrain our imaging AI on new patient data, but only if the new accuracy does not drop below our strictly defined performance floor, and we have triggers in place if the model starts drifting away from its baseline.
Maya: Exactly. Another allowed change is integrating a new input data source or system interface. Think about a new file format, a new data feed, or an interoperability update. The intended purpose has to remain unchanged, and predefined acceptance criteria must confirm performance is maintained.
Sam: That makes total sense for hospital IT teams. Interoperability updates and data integrations are a constant headache, so pre-approving those changes speeds up getting the latest integrations deployed onto the ward without getting bogged down in red tape.
Maya: They also allow for pre-specified updates to an in vitro diagnostic software algorithm parameter set. But again, it must be within defined limits, with fixed clinical decision thresholds, and fully documented verification and validation requirements.
Sam: What about the user interface? Clinicians are always asking for usability tweaks to how a dashboard looks or how patient data is displayed on the screen.
Maya: User interface updates are allowed in the predetermined plan, but only if they do not change the clinical output or the underlying decision logic. And you have to define the usability verification and regression testing entirely in advance.
Sam: And I assume routine maintenance is covered too? Like updating the background systems?
Maya: Yes, planned maintenance updates are permitted. This covers operating system compatibility updates, third party software updates, or what the document explicitly calls "software of unknown provenance". As long as the change boundaries, verification, validation, and criteria are pre-specified and the original intended purpose remains completely unchanged.
Sam: That phrase, software of unknown provenance, is a fascinating technical term. In the software world, that usually refers to off-the-shelf components or open-source libraries that developers pull in without knowing the full, rigorous development history. It is good to see regulators acknowledging that reality.
Maya: It is a practical acknowledgement of how modern software is built. But we need to flip to the other side of the fence now. The document is very explicit about the changes that absolutely cannot be covered by a predetermined change control plan.
Sam: Right, the deal-breakers. What is strictly forbidden from these fast-lane plans?
Maya: The absolute biggest rule is any change to the medical device software's original intended purpose. If your clinical AI is approved to detect a specific lung condition, you cannot use a change plan to suddenly start diagnosing it, or applying that algorithm to a totally different disease.
Sam: Makes sense. You cannot sneak a whole new medical product through a software update.
Maya: You also cannot include broad, open-ended changes. The document specifically calls out vague terms like "future improvements" as completely unacceptable.
Sam: So a medical software developer cannot just write in their application that they plan to make the algorithm better over time, and expect the TGA to just wave it through.
Maya: Definitely not. You are also banned from using these plans for changes to safety-critical attributes. That means you cannot use this pathway to alter clinical decision logic, alarm or alert behaviour, fail-safe behaviour, or other controls relied upon to manage serious risk to the patient.
Sam: Wow, so if a patient monitoring system vendor wants to change the threshold for when an alarm sounds on the ward, they have to go through the standard regulatory pathway. They cannot pre-approve an alarm change using this new framework.
Maya: Correct. Patient alarms are too critical. Furthermore, these plans are strictly for software. Non-software changes, like physical material or hardware changes, must use other regulatory pathways. And finally, you cannot use this for unanticipated changes triggered by safety signals, complaints, cyber security events, corrective actions, or real-world performance issues.
Sam: Wait, really? So if a major cybersecurity vulnerability is discovered in the software, or doctors complain that an AI is missing obvious flags in the real world, the manufacturer cannot use their predetermined plan to push a quick fix?
Maya: Exactly. Because those changes are reactive in nature. They were not explicitly pre-defined within the plan. The document states those reactive issues must be managed through standard change control, post-market compliance actions, or premarket variation applications as necessary.
Sam: That is a crucial distinction for hospital technology buyers to understand. This regulatory fast lane is exclusively for planned evolution, not emergency patching or reactive bug fixes.
Maya: Let us talk about who actually gets to use these plans. The TGA notes they are meant for higher risk-class software. They would normally not be needed for exempt medical device software or for Class I medical device software.
Sam: Is there any exception for the lower risk categories?
Maya: Only if that Class I software has a measuring function. Otherwise, this framework is aimed squarely at higher-risk categories that require active TGA approval.
Sam: The document gives some really practical examples to illustrate how this plays out in the real world. Let us unpack those, because they really connect this regulatory framework to the clinical reality developers are facing.
Maya: The first example involves retraining an algorithm to improve accuracy. A manufacturer has software intended for the measurement and visualisation of the eye, using artificial intelligence to detect and diagnose eye conditions. The intended users are qualified health professionals like optometrists and ophthalmologists.
Sam: Okay, so a highly specialized clinical diagnostic tool.
Maya: Right. The manufacturer wants to use a predetermined plan to improve the sensitivity and specificity of the software using real world evidence collected during the first 2 years of supply in Australia. Because there are no changes to the intended use population or the intended purpose of the device, this is perfectly suited for a predetermined plan.
Sam: And to get that approved, they just have to submit their plans for that real-world monitoring, specify the exact data they will collect, and explain how they will inform users about it. It allows the diagnostic AI to get progressively smarter over its first 2 years without the company filing fresh regulatory paperwork for every single incremental gain.
Maya: The second example shifts from ophthalmology to radiology. This scenario involves an AI-enabled software that interprets and analyses chest x-rays direct from the x-ray system, specifically to highlight areas of concern that a radiologist should further investigate.
Sam: A classic clinical triage and flagging tool for imaging.
Maya: Initially, the software is only compatible with 2 models of x-ray systems. However, the manufacturer plans to do further testing and release a new version in 3 years, which will be compatible with a further 2 models.
Sam: So they use the change control plan to pre-approve that future hardware compatibility expansion. They describe how the new model compatibility will be tested, validated, communicated, labelled, and what user training will be provided.
Maya: Quick note before we carry on. This spot is open for a sponsor. If your company builds or sells AI for healthcare and wants to reach the clinicians, health-system leaders and industry teams who listen to this show, the link to our sponsorship page is in the show notes.
Sam: And now, back to the document.
Maya: Exactly. The intended use population and purpose remain the same, so it fits the rules perfectly. But the third example is where the regulations get really tricky. It is about a mobile phone application.
Sam: I read this part, and it is a fascinating regulatory quirk. The manufacturer launches a mobile application initially with iOS compatibility, but plans to supply an android compatible device within the next 3 years.
Maya: The guidance outlines 2 scenarios that could occur here, depending entirely on the device risk classification. Scenario 1 is if the device is classified as Class III. Under the Australian medical device regulations, Class III software medical devices with different operating systems are considered different kinds of medical device.
Sam: Meaning you cannot use a change control plan to pre-approve the Android version. You have to submit a totally new device application just because it sits on a different mobile operating system.
Maya: Exactly. But in scenario 2, if the device is classified as Class IIb or lower, the two operating systems are considered the same kind of medical device. In that lower-risk case, a predetermined plan can be used to seek premarket approval via a Device Change Request application.
Sam: That is a massive distinction for digital health startups to grasp. Your risk classification dictates whether adding an Android app is a simple pre-planned update or a major regulatory hurdle.
Maya: Now, we need to look closely at the evidence requirements. The TGA is not handing out free passes. A proposed plan must demonstrate to the regulator that the manufacturer is genuinely capable of safely managing the changes.
Sam: The document explicitly warns that proposing an unreasonably high number of changes, or highly complex changes, might reduce the regulator's confidence, and the plan may simply not be accepted.
Maya: Right. Higher-risk products, like Class IIb/III software as a medical device, AI software, software that controls higher-risk devices, and IVD software would normally need much more detailed documentation to build that confidence.
Sam: So what exact details do manufacturers have to provide in the plan to prove they are ready?
Maya: It is highly comprehensive. Under the description of the planned changes, they need to list the objective or reason for the change, any change to device performance, changes to labelling, whether the change will be automatic or manual, and if it affects all models of the device or just one site or region.
Sam: The automatic versus manual distinction is huge for clinical governance. Hospital IT leaders want to know if an algorithm is going to update silently overnight, or if a hospital administrator has to manually accept the new version before clinicians use it.
Maya: Exactly. Then, for the change protocol itself, they must define data requirements and management processes, verification methods, and predefined acceptance criteria that are quantitative, statistically sound, risk appropriate, and clinically meaningful.
Sam: And if the update changes how a clinician interacts with the tool, like a workflow change, they have to include a usability or human factors evaluation appropriate to the risk and the nature of the change.
Maya: Then comes the impact assessment. They must provide a detailed risk assessment demonstrating that the proposed changes will not introduce new, unmitigated risks and that the safety and effectiveness of the device is maintained or improved. They also strictly require a cybersecurity impact assessment.
Sam: Which makes complete sense. If an update changes system interfaces or dependencies, it might open a new backdoor for unauthorized access. The TGA wants to know you have assessed that security risk before the new code is ever written.
Maya: There is a specific section of this document that caught my eye regarding the data used for updates. It focuses heavily on data representativeness or relevance to the Australian population.
Sam: This is critical. If an international technology company trains an AI model on patient data from North America, the TGA demands a solid rationale for why that data remains relevant to Australian settings. They specifically mention demographics, ethnicity, disease spectrum, and local clinical workflows.
Maya: They explicitly require a bias or equity assessment of subgroup performance and bias controls relevant to Australian populations. Manufacturers must justify that retraining datasets remain representative over time.
Sam: That is a powerful safeguard against algorithmic bias creeping in during iterative updates. You cannot just blindly ingest global datasets if it skews diagnostic performance for local patients.
Maya: Alongside that is the requirement for cumulative drift management. The manufacturer must explain exactly how they monitor for cumulative changes to assure the evolving evidence base continues to support the intended purpose and risk profile over the total product lifecycle.
Sam: Right, because an AI might shift slightly with update one, slightly more with update two, and by update ten, it is diagnosing differently than the originally approved version. The plan has to actively monitor that total, long-term drift.
Maya: Transparency is another major pillar. The manufacturer must provide a clear description of how implemented changes are communicated to users, including any necessary training needs and labelling updates.
Sam: And the document states very clearly that planned changes that have not yet been implemented should generally not be described as current device performance. You cannot market a software capability to hospitals just because it is pre-approved in your change plan.
Maya: Manufacturers also have to manage their Unique Device Identifiers carefully. They must determine if the planned change will require a new UDI-Device Identifier or UDI-Production Identifier, based on guidance regarding whether the change meets the threshold for a minor or major change.
Sam: What happens if a company wants to bring in software that already has a change plan approved by a regulator outside of Australia? Like European authorities or other comparable overseas regulators?
Maya: The guidance covers this specifically. A sponsor can apply to include a software medical device on the Australian Register of Therapeutic Goods supported by comparable overseas regulator evidence. They can include a predetermined change plan that was approved by that same regulator.
Sam: But there is a catch to that rule, right?
Maya: Yes. The TGA will not accept a change plan approved by a different comparable overseas regulator than the one used to support the application. The regulatory approvals must match perfectly.
Sam: So if you used one overseas regulator's evidence for the device application, you cannot attach a change plan approved by a different overseas regulator. You have to keep the regulatory pathways aligned.
Maya: Exactly. The sponsor must also provide a redlined comparison, a summary of updates, and a justification for unchanged sections when submitting an updated plan.
Sam: Let us talk about what happens after the software is out there in the real world. Once a manufacturer actually executes a change that was pre-approved, do they just silently push the update to the hospital network?
Maya: When they implement a substantial change covered in an approved plan, they must formally notify the TGA. The regulator will record the notification without further assessment. That is the faster pathway in action, but the notification is still mandatory.
Sam: But the TGA can still audit them after the fact, right? What exact evidence do they have to keep on hand to prove they followed their own rules?
Maya: A substantial amount of evidence. The TGA may request an implementation summary cross-referencing the protocol, the acceptance criteria, and the results. Importantly, this includes documenting failures and exactly how deployment was blocked or remediated.
Sam: That is critical. If a retrained diagnostic model fails its own predefined accuracy tests, the company must document exactly how they stopped it from reaching the hospital network.
Maya: They also require a traceability matrix mapping each implemented modification back to the data sets, acceptance criteria, and impact assessments. They want to see quality management system document control, referencing standards like ISO 14971 and IEC 62304 where applicable. They also expect rollback and contingency arrangements if a deployment resulted in unacceptable performance or safety risk.
Sam: Rollback plans are vital for clinical safety. If a new AI update suddenly starts generating false positives on the ward, the IT team needs absolute confidence the vendor can immediately revert to the stable version.
Maya: The ongoing responsibilities rest heavily on the sponsor. They must maintain sufficient records to determine whether any reported issue or adverse event is associated with a specific software change implemented under a predetermined plan.
Sam: So if there is a patient safety incident, the sponsor has to be able to trace it back to see if an iterative update caused it. They cannot just blame the machine learning process in general.
Maya: Finally, the manufacturer must periodically review the plan itself for ongoing relevance. They should consider updating the plan and re-applying if the underlying software architecture changes materially, or if new regulatory expectations regarding cybersecurity or equity alter the risk assessment criteria.
Sam: This whole document is fundamentally about balancing rapid digital innovation with strict patient safety. It gives developers the speed they desperately want, but demands a level of foresight, rigorous documentation, and ongoing monitoring that will really test their quality management systems.
Maya: To recap, the TGA guidance on Predetermined Change Control Plans allows medical software and artificial intelligence developers to pre-specify strict boundaries for iterative updates, like model retraining or interoperability improvements. If approved, they avoid repeated pre-market assessments.
Sam: But the guardrails are rigid. You absolutely cannot change the intended clinical purpose, you cannot use vague terms, and you must prove that any new training data is relevant to the local population without introducing bias.
Maya: For hospital leaders and IT buyers, the memorable takeaway is that software updates will arrive faster, but you must ensure vendors provide clear communication about exactly what is changing, and verify they have robust rollback plans if a pre-approved update underperforms.
Sam: That wraps up our deep dive into this regulatory shift. As a reminder, you can find a link to the source document in the show notes, and this conversation is not medical advice.